Chapter 3 Assembling
38
8
Attach a tube.
A
ttach the tube to be used to the epi illuminator or epi fluorescence attachment.
The attachment procedure is the same for all tube types.
(1) Completely loosen the M4 set screw for the tube at
the top front of the illuminator using the hex driver
(nominal designation: 2).
Loosening the tube fixing screw
(2) Align and fit the round dovetail at the bottom of the
tube to the round dovetail of the illuminator and
tighten the fixing screw.
P2-TERG100
MADE IN CHINA
Attaching the tube
9
Attach eyepieces.
A
ttach the eyepieces to the tube.
Two eyepieces with the same magnification are used.
Firmly insert the eyepiece until it contacts the end of the
tube sleeve.
BINO
PHOTO
Attaching the eyepiece
Caution when attaching eyepieces
Pay attention when inserting a 10x eyepiece because a rubber cover is attached and the eyepiece end is not visible. To
prevent malfunction, hold the rubber cap when inserting the 10x eyepiece. Do not hold the diopter adjustment ring
when inserting it.