Electrodes
-- +
WHITE RED
BLACK BROWN
BLUE ORANGE
GREEN (Ground)
Electrodes
-- +
WHITE BLACK
WHITE BROWN
BLACK BROWN
Ch. 1
lead I
Ch. 2
mod.
V5
Ch. 3
mod. III
3-Electrode Placement
7-Electrode Placement
Ch. 1
mod.V5
Ch. 2
mod.V1
Ch. 3
lead III
NEMM020 - Rev-I - English
DR200/HE & DR300 Hook-Up Guide for Holter
5-Electrode Placement
Electrodes
-- +
RED BROWN
RED BLACK
WHITE BLACK
GREEN (Ground)
Ch. 1
CM5
Ch. 2
aVF
Ch. 3
lead II
Patient Preparation
• Select area over bone, avoiding muscle areas and breast tissue.
• Clean electrode sites aggressively with alcohol and use a razor to remove hair.
• Abrade electrode sites with a scrub pad or gauze. Allow sites to dry before applying electrodes.
• Firmly snap electrodes to the lead wires.
• Apply electrodes to appropriate sites. Press adhesive border rmly for consistent adhesion.
• To help alleviate tension, loop and tape lead wires down.
Tape
Electrode
TO GET TO MAIN MENU:
• Remove and reinsert battery.
• If 15- second countdown begins,
interrupt by pressing ENTER, ▼, ▲,
and EVENT buttons, in that order.