Post-use Maintenance
6 – 1 Preparation for maintenance
1
Wear eye protection glasses, a mask, and
gloves to prevent infection�
2
Turn off the control unit�
3
Remove the motor handpiece cord plug from
the control unit�
WARNING
Maintain the contra angle head and the file properly as follows� Improper maintenance may cause
infection, product failure, and injuries from overheating:
Be sure to clean the contra angle head and the file immediately (within 30 minutes) after use
to remove residue�
Be sure to disinfect the contra angle head and the file, within 2 hours after cleaning�
CAUTION
Do not clean, immerse, or wipe the product with electrolyzed-oxidizing water (strong acidic water,
super acidic water), strong acidic and strong alkaline liquid solution, solvent containing chlorine,
benzene, or thinner
Follow any additional local directives, standards, and guidelines for cleaning, disinfection,
lubrication and sterilization�
6 – 2 Cleaning and disinfection
[[For contra angle head]]
[Cleaning of Casing]
1
Remove the file from the contra angle head�
2
Remove all debris on the surface of the
product with wipes (Minuten Wipes:ALPRO)�
3
Remove the disposable barrier sleeve from
the contra angle head and the motor
handpiece�
4
Remove the contra angle head from the
motor handpiece�
5
Wipe the debris on the surface of the contra
angle head with a cloth moistened with
water