Post-use Maintenance
CAUTION
Immediately after cleaning, disinfecting and drying cycle is complete, the contra angle head must
be removed from the washer-disinfector (within 1 hour) to prevent corrosion�
After using the washer-disinfector, dry the product completely� Remaining moisture causes internal
corrosion, etc�
After cleaning and disinfecting the contra angle head, be sure to perform “6 3 Lubrication” and
“6 – 4 Sterilization”�
[[For motor handpiece]]
CAUTION
The motor handpiece cannot be cleaned with washer-disinfector
[Cleaning]
1
Wipe the debris on the surface of the motor
handpiece with a cloth moistened with water
2
Under appropriate lighting (
≥500lx
), inspect
the motor handpiece for contamination�
If any visible contamination remains, repeat
the process from procedure No� 1 until it is
visually clean�
500lx
[Disinfection]
1
Wipe the surface of the motor handpiece
with disinfectant wipes (Minuten
Wipes:ALPRO)�