Fig. 7
Coupling Cap
Coupling
Pin
7. Coupling Cap
(
PTL-CL-LED/PTL-CL-LED III
)
9
ENGLISH GERMAN FRENCH SPANISH ITALIAN
8
(
1
)
Connecting
To connect, align and insert carefully and firmly the pin of the coupling cap into the pin
holes of the coupling, and turn the ring until you hear the noise “Click”.
(
1
)
Changing O-Ring
1
)
Remove the worn O-rings.
2
)
Insert new O-rings the O-ring grooves.
KCL-LED
Optional O-rings : Order No. 0318-040100
Optional O-rings : Order No. 0318-068080
SCL-LED
Optional O-rings : Order No. 0312-055100
Fig. 6
Picture shows KCL-LED
6. Changing O-Ring
(
KCL-LED, SCL-LED
)
·Maintenance the coupling after the air supply is stopped completely.
· Do not pull the O-ring with excessive power.
CAUTION
O-ring
(
3 pcs.
)
O-rings
(
2 pcs.
)