4) Keep the components in the autoclave pouch until ready to use.
5) Store in a dry and dust-free location up to the period specified by the sterilization pouch manufacturer. If sterility
cannot be confirmed, sterilize again prior to use.
CAUTION
Do not exceed 138˚C during sterilization.
5. Mounting and Removing the Handpiece
Align the Dots on the Handpiece and the Handpiece Cord. Push
handpiece into connector.
To detach the handpiece, grip the Handpiece and plug of the
handpiece cord and separate in a straight motion.
(
Fig. 7
)
6. Mounting and Removing Tip
Fig. 8
Loosen
Tighten
Tip Wrench
Tighten
Loosen
Turn TIP lightly by hand, and install it.
Tip will insert from the bottom hole of Tip Wrench. Align the four corner of the Tip base area into the four corner of Tip
Wrench. And turn it clockwise until it clicks.
Do not touch the top part of TIP to avoid an injury.
To remove the Tip, turn it counterclockwise with the Tip Wrench.
Dots
Handpiece backend
Handpiece Handpiece Cord
Fig. 7
CAUTION
Always confirm that the handpiece is correctly seated and locked into place.
Do not use a Handpiece other than that supplied with this device.
WARNING
To avoid Electrical Shock Do not touch the handpiece backend electrical contacts.