* Operation Principle
A sinusoidal electrical signal, at ultrasonic frequency
(
f
>
20kHz
)
, is delivered by a generator. This signal is applied
to the ‘piezoelectric ceramic’ located inside the transducer. Piezoelectric ceramic converts this signal into mechanical
vibrations. These vibrations are at the same ultrasonic frequency as the electrical signal. The mechanical vibrations
are propagated towards the distal end of the transducer. The “TIP” insert, which is attached at the distal end of the
transducer, vibrates at ultrasonic frequencies and makes it possible to achieve the aimed purpose.
· During operation, high frequency oscillations in the handpiece and handpiece cord may affect computer and LAN Noise
may be heard during operation near a radio receiver.
· Be sure to turn off the Power/Volume Knob after use. Remove the AC Adaptor when not using it for long periods of time.
·
The surface of the Control Unit may feel warm after prolonged use. Turn the Power/Volume Knob to OFF and cool the device.
· Users are responsible for operational control, maintenance and inspection.
· Clean/sterilize the product immediately after using it. Then store it. Leaving it non-sterile might lead to failure.
· When you have not used the product for long time and use it again, check the operation before use.
· When abnormalities are found with a Control Unit and /or an AC adaptor, pull the AC adaptor from AC Outlet immediately.
· Do not turn the Power/Volume Knob to ON or OFF needlessly.
· No special training is required for this device.
· Applied parts for patient and/or operator are/is Tip and Handpiece.
· When using a file as a tip for root canal treatment (for enlarging, cleaning, etc.), use it after confirming its characteristics
and areas where it can be applied.
· Operators and all others in the area must wear eye protection and a mask when operating this product.
· Before use, be sure to always check that the power level is set to the proper settings.
· During vibration, do not grasp the tip nor make contact with the tip for an extended time.
· Do not use tips that are damaged, bent, corroded, or show other imperfections. The tip could break or fail to vibrate.
· Do not apply more pressure than required to the tip end. This could cause the tip to break or warp. Do not twist or strain
the tip, and do not use it as a replacement for surgical tools. This could cause the tip to deform or break.
· If the tip cracks or breaks during use, immediately stop operation, and replace with a new tip. Operation with a damaged
tip could injure the gums, periodontal tissue, or other locations.
· The diamond coating on the tip surface may peel off depending on the tooth condition or root canal condition. If the
diamond coating peels off, the cutting efficiency becomes less, resulting in reduced cutting performance. If you sense this
type of condition, replace with a new tip.
· All sterilizable components of this device are delivered in a non-sterile condition and must be autoclaved prior to use. Refer
to "4.Sterilization before first time use".
· The diamond coated tips are "Single use". Do not reuse.
· U.S. Federal law restricts this device to sale by or on the order of a licensed physician.