- Page 38 -
Nu-Tek Levator Elite User Manual
Guidance and manufacturer’s declaration - electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment - guidance
The Nu-Tek Levator Elite is intended for use in the electromagnetic environment
specied below. The customer or the user of the device should assure that it is used
in such an environment.
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all
establishments , including domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
RF emissions
CISPR 11
RF emissions
CISPR11
Harmonic
emissions
lEC 61000-3-2
Voltage
uctuations /
icker emissions
lEC 61000-3-3
Group 1
Class B
Not
Applicable-
Battery
Operated
Device
Not
Applicable-
Battery
Operated
Device
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
The Nu-Tek Levator Elite is intended for use in the electromagnetic environment
specied below. The customer or the user of the device should assure that it is used
in such an environment.
Floors should be wood, concrete
or ceramic tile. If oors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
Electrostatic
discharge (ESD)
lEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
±6 kV contact
±8 kV air
±2 kV for
power supply
lines
±1 kV line(s)
to line(s)
±6 kV contact
±8 kV air
Not Applicable-
Battery
Operated
Device
Not Applicable
- Battery
Table 1
Table 2