ALARM REASON SOLUTION
The pump has
administered the dose
Press and hold the ON/OFF
button until it beeps and is
powered off
If the END OF DOSE alarms
inappropriately contact your
Nutricia Homeward Enteral Nurse
Specialist or the Out of Hours
Advice Line
The pump detected an
occlusion between the
feed and the pump
Make sure there are no kinks in
the tubing, that clamps are not
closed between the feed container
and the pump, and that the foil has
been pierced
The sensor is dirty
Clean the sensor area (see
cleaning guidelines), reinsert the
giving set in the pump and restart
the pump
The pump detected an
occlusion between the
pump and the patient
If relevant, make sure the clamp
is open
Make sure there are no kinks in
the giving set or feeding tube
The sensor is dirty
Clean the sensor area (see
cleaning guidelines), reinsert the
giving set in the pump and restart
the pump
Please refer to the step by step guide on next page è
CAUTION: Do not use the ‘FILL SET’ funcon when there is a set occlusion
as this may cause the pump to incorrectly calibrate
OCC OUT
OCC IN
END OF
DOSE