Flocare Infi nity+
Problem solving
instructions
Error messages and instructions
Please refer to detailed Instructions for Use for additional information
171101 / BP&MC 11090802
> Press to switch off the pump > Insert the Flocare
®
Infi nity
™
set into the pump following
instructions for use > Restart the pump.
> The pump has been untouched in hold mode for over 3 minutes > Extend the hold mode by a further
3 minutes by pressing > Program the pump and press to start the pump.
> The programmed dose / volume has been administered > Switch off the pump by pressing
for 2 seconds > Clear the memory by pressing .
> No fl ow rate is installed > Make sure the correct fl ow rate is programmed.
> The door is not correctly closed > Make sure the door is properly closed prior to starting a feeding
program.
> Check if the set is inserted correctly and check the air sensor > Remove the air in the set by using
key.
> Connect the adaptor to the pump and mains and charge the pump for approx. 6 hours. During charging
the pump can be used.
> Battery failure > Please refer the pump for service.
> Press > Check the feeding set (IN) or feeding tube (OUT) for blockages > Reinsert
the set and start the pump, if necessary clean the sensors.
> Electronic error > Switch the pump off, make sure the pump door is closed and turn the pump back
on > If the error persists, please refer the pump for service.
> The wall outlet or the adaptor does not work > Connect to another wall outlet > If there is still no
symbol, request replacement of the adaptor.
NO SET
PUSH STRT
END OF DOSE
PROG
DOOR
AIR
BATT
BATT E + F
OCC IN
> The programming feature is blocked in the set-up mode of the pump > Refer to instruction booklet
to de-block.
LOCK
ER01 - ER99
NO PLUG SYMBOL
OCC OUT
Problem Instructions
(while the pump is
connected to the mains)
clr
fi ll set
start/stop start/stop
start/stop
> Stop the pump by pressing the > Let it run briefl y by repressing the
key > Stop the pump again with key, ensuring there has been no OCC OUT alarm >
Remove the set from the pump and reinsert it into the pump > Restart the pump by pressing
.
OCC OUT
(repeated)
start/stop start/stop
start/stop
start/stop