EN-10 RS232 Interface
For PJX balances:
Command
Function
Immediate Print of displayed weight (stable or unstable).
Print displayed weight (stable or unstable).
Set Auto Print to On Stability, allow non-zero displayed weight be printed.
Set Auto Print to On Stability, allow both stable non-zero weight and stable zero reading printed.
Set Auto Print to Interval Print, x = print interval (1-3600 sec), 0P disable the interval Print
0P disable interval print, continuous print or print on stability
Enter Print Header Lines, the format is: H x "header string".
Where x = line number 1 to 3, “header string” can be up to 24 alphanumeric characters.
If no string in the command, "H x" will read the stored header x.
Same as pressing Zero Key.
Same as pressing Tare Key.
Establish a preset Tare value in displayed unit. x = preset tare value. Sending 0T clears tare (if allowed).
Prints Tare weight stored in memory.
Print current application mode (weighing mode).
Set current application mode to x. x depends on applications
1 – Weigh
2 – Count
3 – Percent
4 – Density
Scroll to the next enabled mode.
Print Current weighing unit: g, kg, lb, oz, etc….
Set balance to unit x: g, kg etc.
1 – g
2 – kg
3 – mg
4 – ct
…
Scroll to the next enabled unit.
Begin Internal Calibration.
Print Version: print name, software revision and LFT ON (if LFT is set ON).
Set Counting APW (x) in grams. (must have APW stored)
Print Counting application APW.
Set Percent application reference weight (x) in grams. (must have reference weight stored)
Print Percent application reference weight.
Set Time x format: hh:mm:ss
Set Date x format: mm/dd/yyyy
0 = print unstable data, 1 = print stable only
Enable or disable OK response to non-print commands: x=0 to disable, x=1 to enable.
Pre-tare the container weight (x) in grams.