S
Samuel PayneAug 14, 2025
What to do if Ohio Medical Corporation MiniOX 3000 displays 'ERR' and an error code?
- SSara GilesAug 14, 2025
If the Ohio Medical Corporation Medical Equipment displays 'ERR' along with an error code (01 to 08), accompanied by visual and audible alarms, and the monitor locks up, it indicates the microprocessor has detected an internal error. Note the error code and consult Appendix E, Error Codes. You can also try disconnecting the battery. If the issue persists, contact Ohio Medical Corporation service.