61
Appendix
. Technical Data on CASY TT
Function: Electronic Cell Counter + Analyzer System for determining cell counts
and cell size distributions, aggregation and viability.
Operating Environment: Indoor use only
Operational altitude: Up to 2000 m above sea level
Dimensions: Height: 39 cm
Width: 31 cm
Depth: 39 cm
Weight: 19 kg (including control panel)
Ambient conditions: Maximum operating temperature:
Recommended operating temperature:
Operating humidity:
10°C to 40°C
15°C to 32°C
30% to 80% relative
humidity
(non condensing)
Power supply: 230 V ~ +/- 10% / 50 Hz
115 V ~ +/- 10% / 60 Hz
Power consumption: 60 VA
Fuses (primary): 230V: 2 × 800 mAT
115V: 2 × 1.6 AT
Fuses (secondary): SI1: T 160mA
SI2: T 1.6A
SI3: T 400mA
SI4: T 2.5A
SI5: T 400mA
Installation category: II (according to IEC 61010-1)
Volume measuring section and
pressure system:
Calibrated, opto-electronically controlled volume measuring tube
with no-maintenance, sensor controlled pressure system. Mercury-
free.
Measuring principle: Electrical resistance measurement in 1 MHz, low voltage eld with
pulse area analysis by Schärfe System.
Distinctive measuring channels: > 512.000.
Volume resolution: 1 in 512,000
Displayed size channels: 400
Measuring range: 0.7 µm to 120 µm
Measurement dynamics: Ratio of smallest to largest particles that can be analysed
simultaneously:
>1 : 70,000 in terms of volume, >1 : 40 in terms of diameter
Typical analysis time: 10 sec
Typical sample volume: 5 µl – 100 µl
Measuring volume: 200 µl – 4 ml
Pollution degree: 2 (according to IEC 61010-1)
Technical Data on CASY TT