IMPACT ECHO (IE)
www.olsoninstruments.com www.olsonengineering.com
# Recs. Number of Record is a total number of IE data records you want to save. This parameter
can not be changed after the data was taken.
Pre-Trigger is number of points before the triggering point that data collection starts. This
parameter must be 0 for the IE test. This parameter can not be changed after the data was
taken.
Trigger Level % is the minimum signal amplitude (in terms of percentage full scale) to trigger
data acquisition. In this case, the trigger level was set at 6%. Therefore, the system will start
acquiring data once the absolute amplitude of the signal exceeds 0.6 volt. This parameter can not
be changed after the data was taken.
Channel Setup. For the IE test, only one channel (connected to the IE head) will be used. Ch 1
must be turned on for the IE test and this cannot be changed.
Sol. This value can be “on” or “off”. If the solenoid is on, the system will use the solenoid in the
IE head as a source. If this value is off, the system will wait for the external hammer as a source.
Sol Time. This is the time to drive the solenoid. The default value is 12 us. However, this value
may need to be higher for the underside or underwater testing.