IMPACT ECHO (IE)
www.olsoninstruments.com www.olsonengineering.com
7. If the solenoid mode is on, the NDE-360 system will fire the solenoid inside the IE head and
the displacement transducer in the IE head will record the data. If the solenoid mode is off,
hit the external hammer on the structure next to the IE head (or accelerometer). The screen
below shows the spectrum of the time domain data acquired from a 5 in concrete cylinder.
Press “A” button
8. Repeat Step 7 until all the records are collected.
9. After all the records are collected. Press “2” to save the data or “1” to exit without saving.
10. To perform more IE tests, repeat Steps 7 – 9.
11. The data can be recalled in the NDE-360 system by pressing the “File” button on the main
screen and then select Option 2.
12. The data files on the compact flash card can be moved to the analysis computer for post-data
analysis.