DMTA-10055-01EN, Rev. A, February 2015
Calibration
159
Figure 9-11 Entering the Zero Cal thickness value
11. Adjust the value to match the known thickness of the gated indication (0.100 in. in
this example), and then choose Continue to continue to the second calibration
step. The thickness value used in this part of the calibration will be stored for
reference in the Cal-Zero parameter box.
To exit without acquiring calibration data, press Cancel.