DMTA-10026-01EN, Rev. D, January 2015
Chapter 5
92
The units and resolution of the differential thickness are the same as those selected for
the thickness measurement.
When you press [SAVE] while in differential modes, the Magna-Mike 8600 saves both
the live and differential measurement values. The user can select to display the
differential value or the live thickness measurement in large font. This is selected in
the DISPLAY menu.
To activate and configure a differential mode
1. In the measurement screen, press [SET UP], and then highlight DIFFERENTIAL.
2. In the DIFFERENTIAL submenu, select ENABLE.
3. In the DIFFERENTIAL SETUP screen (see Figure 5-2 on page 93):
a) Set ENABLE to ON to activate the differential function.
b) In DIFF TYPE, select one of the two differential types:
NORMAL: Shows the actual thickness along with the difference between
the actual thickness measurement and the REF VALUE that you entered.
PERCENT: Shows the actual thickness along with the percent difference
from the REF VALUE that you entered.
c) In REF VALUE field, set the desired reference value.
Differential
Normal
Current thickness Reference value=
Differential
% Ratio
Current thickness Reference value
Reference value
---------------------------------------------------------------------------------------
100×=