2. In TARGET SELECT, when using 86PR-1 and 86PR-2 probes, set to Auto Ball for
the standard target diameter balls:
1.59 mm (1/16 in.), Olympus P/N: 80TB1
3.18 mm (1/8 in.), Olympus P/N: 80TB2
4.76 mm (3/16 in.), Olympus P/N: 80TB3
6.35 mm (1/4 in.), Olympus P/N: 80TB4
4.76 mm (3/16 in.) magnetic, Olympus P/N: 86TBM3
6.35 mm (1/4 in.) magnetic, Olympus P/N: 86TBM4
When using the 86PR-3 probe, the Magna-Mike 8600 is unable to automatically
recognize targets, so for this type of probe you must manually select the 1.59 mm
(1/16 in.) ball, 3.18 mm (1/8 in.) ball, 0.66 mm (0.026 in.) wire, or target disk in
TARGET SELECT.
The Magna-Mike 8600 is unable to automatically recognize a target disk or wire. You
need to manually select the following targets:
.5 Disk for 12.70 mm (0.500 in.) diameter disk, Olympus P/N: 80TD1,
V Disk for 6.35 mm (1/4 in.) diameter V disk, Olympus P/N: 80TD2,
0.026 Wire for 0.66 mm (0.026 in.) diameter target wire, Olympus P/N: 86TW2,
0.045 Wire for 1.14 mm (0.045 in.) diameter target wire, Olympus P/N: 86TW1.
3. In WEARTIP SELECT, select the type of wear tip that is being used. Most
commonly and by default, this will be set to STANDARD (for the standard wear
tip and the extended wear tip). The user must manually select CHISEL when a
chisel wear tip is being used. The measurement screen of the Magna-Mike 8600
shows an icon that indicates what the current wear tip selection is:
A white dot next to the probe type indicates that the standard wear tip is
selected.
A white triangle next to the probe type indicates that the chisel wear tip is
selected.
4. In UNITS, select between the ENGLISH (inches) or METRIC (millimeters) units.
5. In RESOLUTION, select one of the following resolutions:
AUTO (default): Automatically uses HIGH resolution for measurements below
4.06 mm (0.160 in.) and STANDARD resolution for measurements above
4.06 mm (0.160 in.). For the 1.59 mm (1/16 in.) ball, the change occurs at 2.03 mm
(0.080 in.) instead of 4.06 mm (0.160 in.).
STANDARD (0.01 mm [0.001 in.])