DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
67
indicated by a white dot at the top of the screen. The user needs to select the CHISEL
tip in the MEASUREMENT SETUP screen (see Figure 4-4 on page 67) when using the
chisel wear tip, indicated by a white triangle at the top of the screen.
Figure 4-4 The MEASUREMENT SETUP screen
The wear tip should only be hand tightened. Using tools to tighten the wear tip can
permanently damage the probe.
Do not use the Magna-Mike 8600 probe without a wear tip. Doing so will cause
inaccurate measurements and may cause permanent damage to the probe.
4.5 Proper Target Selection
The Magna-Mike 8600 can be used with a variety of target balls of different sizes,
target disks, and a target wire. Choosing the proper ball will greatly improve the
accuracy of the measurements. Use only Olympus target balls, disks and wires with
the Magna-Mike 8600, as other seemingly identical targets may cause inaccurate
measurements.
CAUTION