DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
73
Table 7 Calibration kits
Part number Description
Targets included
in kit
Items included in kit
86ACC-KIT
(U8771068)
Standard calibration kit
for most applications
where thicknesses are
up to 7.62 mm
(0.300 in.).
1.59 mm (1/16 in.),
3.18 mm (1/8 in.),
and 4.76 mm
(3/16 in.) steel
target balls
80CAL-TB1,
80CAL-TB2,
80CAL-TB3,
80TB1, 80TB2, 80TB3,
80CAL-010, 80CAL-020,
80CAL-040, 80CAL-080,
80CAL-160, 80CAL-240,
and 80CAL-300 in a
plastic case.
86ACC-ER-KIT
(U8771069)
Extended range
calibration kit for
applications where
thicknesses are
between 1.02 mm
(0.040 in.) in to
25.40 mm (1.00 in.).
6.35 mm (1/4 in.)
steel balls,
4.76 mm (3/16 in.)
magnetic balls,
and 6.35 mm
(1/4 in.) magnetic
balls
80CAL-TB3,
80CAL-TB4, 86TBM3,
80TB4, 86TBM4,
80CAL-040, 80CAL-160,
80CAL-240, 80CAL-360,
86CAL-500, 86CAL-750,
86CAL-875, and
86CAL-1000 in a plastic
case.
86ACC-D-KIT
(U8771071)
Disk calibration kit for
applications requiring
target disks
12.70 mm
(0.500 in.)
diameter flat disk
and 6.35 mm
(0.250 in.)
diameter V-Edge
disk
86PR1-CWC,
86CAL-TD, 80TD1,
80TD2, 86DCAL-010,
86DCAL-020,
86DCAL-040,
86DCAL-080,
86DCAL-160,
86DCAL-240, and
86DCAL-360 in a plastic
case.

Table of Contents