DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
79
Place the target ball into the BALL ON FIXTURE before placing the fixture onto the
probe. Dropping the ball repeatedly into the fixture as it rests on the probe may pit the
end of the wear tip, causing small measurement inaccuracies.
Figure 4-6 Alignment of target ball in fixture on probe
When using the target disk and the chisel wear tip, make sure that the disk is aligned
perpendicularly to the chisel edge (see Figure 4-7 on page 80).
NOTE
Targ et ball
Probe
BALL ON FIXTURE
IMPORTANT