DMTA-10026-01EN, Rev. D, January 2015
Basic and Multipoint Calibration
83
4.7.3 Saving and Recalling a Calibration File
The Magna-Mike 8600 automatically saves the last calibration that was performed for
each target ball, disk, or wire. The calibration file is automatically named, based on
the target that was used, as shown in Table 10 on page 83.
Each time a new calibration is performed, the Magna-Mike 8600 will automatically
replace the previous calibration file in the instrument memory. This allows the user to
quickly change between the last calibrations performed with different target sizes. If
the probe is changed or if the wear tip is replaced, the user should perform a new
calibration.
To recall a calibration file
1. Press [FILE], and then select CAL RECALL.
Table 10 Calibration file names
Calibration File name Description
1/16 BALL 1.59 mm (1/16 in.) steel target ball
1/8 BALL 3.18 mm (1/8 in.) steel target ball
3/16 BALL 4.76 mm (3/16 in.) steel target ball
1/4 BALL 6.35 mm (1/4 in.) steel target ball
3/16 MAGBALL 4.76 mm (3/16 in.) magnetic target ball
1/4 MAGBALL 6.35 mm (1/4 in.) magnetic target ball
0.50 DISK 12.70 mm (0.500 in.) diameter flat disk
V-DISK 6.35 mm (0.250 in.) diameter V-edge disk
0.045 WIRE 1.14 mm (0.045 in.) diameter wire target
0.026 WIRE 0.66 mm (0.026 in.) diameter wire target
TIP