5. MEASURING METHOD
(1) Turn on the power to the control box (UCB-01).
(2) Turn on the power to the light source power supply (ULP-01).
(3) Turn on the power to the personal computer.
* For details, refer to "USPM-RU Software Operation Manual."
(4) Double-click the USPM icon on the computer desktop to start the program.
*If the control box power is off or the USB cable is not connected, the communication
error message "The communication encountered a device error" is displayed. At
this time, click [OK] to open the menu window, and choose [File (F)]-[Quit program
(X)] in the File menu to quit the program.
The message "Save the setting?" is displayed. Click [NO (N)].
As you return to the desktop, turn on the control box, and double-click the USPM icon
again to launch the program.
(5) After turning on the power to the component parts, give them 15 minutes of
warming-up.
(6) Press the Back Ground button on the measurement screen to conduct
background measurement.
(7) Look into the eyepiece, and adjust the diopter to bring the cross wire into focus. For
this diopter adjustment, turn the helicoid of the scale (-5 to +5) of the eyepiece.
When the computer is misbehaving or any failure is found
in the measurement data, plug off and reinsert USB cable
and then reboot the software.
If you still have such troubles after rebooting the software,
please turn off the Control Power Supply Box and then
reboot the computer.
Even if you still have such trouble after rebooting the
computer, please contact to us.