Sterilization
Drying time
For steam sterilization, Olympus recommends a drying time of at
least 15 to 30 minutes.
Choose a suitable drying time, depending on the autoclave.
Refer to the instructions for use of the autoclave.
9.2.3 After sterilization
Removing from the autoclave
1. Remove the product from the autoclave.
2. Let the product cool down at room temperature.
Do not use additional cooling.
3. Check that the sterilization wraps or pouches are not
damaged.
4. Resterilize the product if 1 of the following applies:
Sealing is open
Packaging is wet