Appendices
C-ii
Low Range -13mV to 33 mV
-13 mV
0.005 mV
0.00194 mV
-10 mV
0.005 mV
0.00192 mV
0 mV
0.0046 mV
0.0019 mV
10 mV
0.0049 mV
0.00192 mV
30 mV
0.00545 mV
0.0021 mV
33 mV
0.00545 mV
0.00214 mV
CHANNEL
STANDARD
MEASUREMENT RESULT
LIMITS OF ERROR
EXPANDED
2
UNCERTAINTY
PASS/FAIL
L
OWER
LIMIT
U
PPER
LIMIT
Cold Junction Compensation
Note: The values entered in the “Standard” column below are the Reference Thermometer readings for each
channel. Calculate your Limits of Error by: first add a factory determined offset of 0.06°C to the “Standard” value,
then add/subtract 0.11°C from the result to determine upper/lower limits. See section 4.2.1 paragraph 15 for an
example calculation.
1
°C
°C
2
°C
°C
1
The estimated expanded uncertainties are based on a standard uncertainty multiplied by a coverage factor K=2,
providing a level of confidence of approximately 95%
2
Published temperature uncertainty values can only be achieved if the temperature of UUT is known within 40mK.
Customers performing CJC adjustments in their facility will need to calculate their own CJC uncertainty.