Adjustment key of thermocouple 0°C
: Main display shows (T1 or T2) readings, use and main display
(T1 or T2) thermocouple of the same type,
imports input of main display 0°C (32°F) of
standard temperature value, adjust the cold
junction compensation of third display, enables
main display to show 0°C (32°F), then press
key finish adjusting. The adjusting range is about
0°C ~50°C (32°F~122°F) of third display cold
junction compensation, exceed range show Err
and unable to adjust, need to press key to
g
ive up and left. The cold junction compensation
adjusting (cancelled) in power off, while starting
the meter again will recover the normal cold
junction compensation.
Press once: Enter adjust manually and third
display shows the cold junction compensation,
enable main display shows 0°C (32°F).
>
Press to move right one-figure that adjusted.
>Press to move left one-figure that adjusted.
>
Press to add one-figure upward that adjusted.
>Press to minus one downward that adjusted.
Press for 2 seconds: Adjust once automatically the
cold junction compensation of the third display,
Make main display shows 0°C (32°F). If need to
adjust again, enter again after press key to
leave.
ADJ
TC O C
C
F
MAX
IN
ADJ
TC O C
ADJ
TC O C
ADJ
TC O C
6
TC O °C
ADJ
TC O °C
ADJ
TC O °C
ADJ
°C
°F
MAX
MIN