3
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable.
Measurement Range:
Thermocouple Range
K-TYPE(0.1°C)
-200°C to 1372°C, -328°F to 1999°F
J-TYPE(0.1°C)
-210°C to 1200°C, -346°F to 1999°F
T-TYPE(0.1°C)
-200°C to 400°C, -328°F to 752°F
E-TYPE(0.1°C)
-220°C to 1000°C, -364°F to 1832°F
S-TYPE(1°C)
0°C to 1767°C, 32°F to 1999°F
Auto range: 0.1°C/1°C, 0.1°F/1°F.
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
±(0.1% rdg + 1°C) range -60°C to 1372°C
±(0.1% rdg + 2°C) range -60°C to -220°C
±(0.1% rdg + 2°F) range -76°F to 1999°F
±(0.1% rdg + 4°F) range -76°F to -364°F
S-TYPE ±(0.0% rdg + 2°C) on 0°C to 50°C
±(0.1% rdg + 1°C) on 50°C to 1767°C
±(0.0% rdg + 4°F) on 32°F to 122°F
±(0.1% rdg + 2°F) on 122°F to 1999°F