2-6     General Information & Specifications 
878595  Personal Daq User’s Manual 
At a measurement duration of 610 milliseconds, the last two bits are considered unreliable, resulting in a 
resolution of 22 bits.  At a very fast measurement duration (12.5 milliseconds), the nine most least 
significant bits are unreliable, resulting in 15 bit accuracy. 
Speed vs. Resolution
1
 Speed Designation 
Measurement 
Duration 
(per channel) 
Maximum 
Sample Rate
2 
(Samples/sec) 
Resolution
2
(Bits RMS) 
(-4 V to +4 V range) 
Very Slow, 50 / 60 Hz rejection  610 ms  1.6 / sec  22 
Slow, 50 Hz rejection  370 ms  2.7 / sec  22 
Slow, 60 Hz rejection  310 ms  3.2 / sec  22 
Medium, 50 Hz rejection  130 ms  7.7 / sec  21 
Medium, 60 Hz rejection  110 ms  9.2 / sec  21 
Medium  40 ms  25 / sec  19 
Fast  20 ms  48 / sec  17 
Very Fast  12.5 ms  80 / sec  15 
Notes: 
1. Each channel can have independent measurement duration and resolution. 
2. The sample rates and resolutions shown were obtained with a 10-channel scan and with 
continuous self-calibration disabled. 
3. Duration does not include the use of CJC measurements. 
 
Note:  When measuring variable input signals (as opposed to relatively steady input signals), the variable 
signals will require more samples/sec to obtain a realistic signal representation.  Available sample 
rates range from 1.6 samples per second up to 80 samples per second as indicated in the preceding 
table. 