EasyManua.ls Logo

Omron HEM-9000AI - Measurement Preparation: Cuff Application; 9.1. Applying the Blood Pressure Cuff

Omron HEM-9000AI
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
9.Measurement Preparation
9. Measurement Preparation
9.1. How to Apply the Blood Pressure Cuff
1. Select a cuff according to the circumference of the
patient’s arm. [Fig. 1]
* Use an appropriate cuff according to the circumference
of the patient’s arm.
* If an inappropriate cuff size is used, the device might
not measure blood pressure accurately.
* Check before wrapping the cuff that:
1. The bladder is set inside the cuff.
2. The tube is not twisted in the cuff.
3. The tube of the bladder emerges from the cuff as
shown in the figure at right. [Fig. 2]
2. Connect the air tube firmly.
* When connecting the air tube to the main unit, connect
the air tube plug firmly into the air connector as shown
in [Fig. 3].
* Connect the air tube and the cuff / bladder set by turning
as shown in [Fig. 4].
* A total air tube length of 1.5 m is provided with the Air
Tube (HEM-9T1.3) in the package.
Arm circumference Name of the arm cuff
17 - 22cm Small size adult arm cuff
22 - 32cm Medium size adult arm cuff
32 - 42cm Large size adult arm cuff
INDEX
ART.
RANGEMIN
MAX
32
~
42cm
(
13"
~
17"
)
REF
HEM
-
CL19
RANGE
MIN
MAX
INDEX
ART.
22
~
32cm
(
9"
~
13"
)
REF
HEM
-
CR19
ART.
RANGEMIN
MAX
INDEX
REF
HEM
-
CS19
17
~
22cm
(
7"
~
9"
)
TUBE
TUBE
TUBE
ART.
TUBE
TUBE
Air Connector
Air tube plug
Air tube sideCuff side
Closeup of the O part
[Fig. 1]
[Fig. 2]
[Fig. 3]
Air tube
Cuff
[Fig. 4]
Where the air
tube emerges
Large size adult arm cuff
Medium size adult arm cuff
Small size adult arm cuff

Table of Contents

Related product manuals