38
13.Blood Pressure Measurement
13.2. Auscultation
When selecting blood pressure measurement, it is possible to take the reading using auscultation.
1. Apply the blood pressure cuff around the patient’s
arm. [Fig. 1]
2. Press the F2 key [AUTO MANU] on the blood
pressure measurement screen to bring up
[MANUAL] on the display. [Figs. 2 and 3]
Check that the measurement method display in the
lower-left part of the screen is set to [MANUAL].
3. Place the stethoscope on the patient’s arm.
4. Press the START button. Inflation begins. [Fig. 4]
* Do not press the START button without applying the cuff.
* When you want to stop during measurement, press the STOP
button. The cuff rapidly deflates.
Operator - controlled (Manual) Inflation.
* Holding the START button down during inflation switches to
manual inflation. During manual inflation, inflation only takes
place while the START button is pressed.
[Fig. 2]
[Fig. 3]
[Fig. 4]
[Fig. 1]