 0HDVXUHGYDOXHVUHÀHFWLQWHUQDOGDWDREWDLQHGE\1(;7*(1(5$7,21,03$&7251*,DFFRUGLQJWR
EN13544-1:2007 + A1:2009.
** The Nebulization rate is measured with saline 0.9% solution at 23°C at spray angle of 30° with 3 ml of
medication. It can vary with medication and ambient conditions.
0
10
20
30
40
50
60
70
80
90
100
0.1 1 10 100
Particle Size Dp (µm)
Cumulative % particle mass of sodium fluoride under size
Individual
Average
MMAD 4.5 μm
Cumulative Undersize %
Durable period Durable periods are as follows, provided the product is used to nebulize
a medication twice a day for 10 minutes each time at room temperature
(23°C).
Durable period may vary depending on usage environment.
Frequent use of the product may shorten the durable period.
Main Unit 5 years
AC Adapter (optional) 5 years
Medication Container 1 year
Mesh Cap 1 year
Mouthpiece 1 year
Mask (PVC) 1 year
Mask Adapter 1 year
 6SHFL¿FDWLRQV