24
ERROR MESSAGES AND DETAILS ABOUT YOUR SYSTEM
Error messages and details about your system
Understanding error and other messages
The OneTouch® UltraMini® Meter displays messages when there are problems with the test strip, with the meter, or when your
blood glucose levels are higher than 33.3 mmol/L or lower than 1.1 mmol/L. Messages do not appear in all cases when a problem
has occurred. Improper use may cause an inaccurate result without producing a warning message.
Message What it means What to do
You may have a very low blood glucose
level (severe hypoglycemia), lower
than 1.1 mmol/L.
This may require immediate
treatment according to your
healthcare professional’s
recommendations. Although this
message could be due to a test error,
it is safer to treat rst and then do
another test.
You may have a very high blood
glucose level (severe hyperglycemia),
over 33.3 mmol/L.
Re-check your glucose level. If the
result is HI again, obtain and follow
instructions from your healthcare
professional without delay.
The meter has detected that the
temperature is above the system
operating range. Do Not perform a
test until the meter and test strips
reach a temperature within the
operating range of 6–44°C.
Repeat the test after the meter and
test strips have reached a temperature
within the operating range.
The meter has detected that the
temperature is below the system
operating range. Do Not perform a
test until the meter and test strips
reach a temperature within the
operating range of 6–44°C.
Repeat the test after the meter and
test strips have reached a temperature
within the operating range.