Take a test2
66
Out-of-range results may be due to:
• Not following the instructions beginning on
page61.
• Control solution is contaminated, expired, or past
its discard date.
• Test strip or test strip vial is damaged, expired, or
past its discard date.
• Meter, test strips and/or control solution were
not all at the same temperature when the control
solution test was performed.
• A problem with the meter.
• Dirt or contamination in the small well on the top of
the control solution cap (see Step 2).
After getting a control solution result
Once you have a control solution result, you may:
• Press and hold
to return to the Main Menu.
Or,
• Press and hold
for several seconds until the
meter turns off. The meter will also automatically
turn off if left alone for two minutes.