STEP 2 Setup the battery
Slide in both batteries into the tray.
Make sure that the battery is being slid-in at the right direction.
INSTRUCTIONS FOR USING THE BATTERIES IN VENUS-series
The Venus batteries are based on LI-ION Technology. A feature of this technology is that if the batteries are discharged until a certain point,
they cannot be recharged again and internally the battery is destroyed. This is inherent of the LI-ION technology.
In order to avoid that Venus batteries are not chargeable anymore we advise to use the Venus batteries as follows. If customers don’t follow
these instructions than the warranty on batteries is void.
1. If you have a failing battery, send the battery immediately to the nearest service partner. If you fail to do this the warranty is void.
2. If a battery is empty, charge the battery immediately.
3. If batteries are not charged immediately the following discharge table will apply. If batteries reach this point they are destroyed and
cannot be charged again. The following table shows in 2 different situations the number of days before the battery is destroyed.
Battery remaining
capacity percentage
Approximately days to over-discharge
(not in VENUS)
Approximately days to over-discharge
(put in close down VENUS-series)
100%
1000 days
130 days
75%
750 days
105 days
50%
500 days
80 days
25%
250 days
55 days
10%
100 days
40 days
5%
50 days
35 days
0%
30 days
30 days