Unscrew the FC/APC connector from the source bulkhead and repolish as needed.
Once the connector has been repaired, turn the unit on and navigate to Scan mode
and verify that there is no noticeable reflection at the front panel.
After checking for reflections, mate the FC/APC connector back to the source panel
bulkhead. Next, slowly replace the panel back on the front of the unit, making sure
that the fiber slides back inside of the unit.
Finally, fasten the source panel to the front of the unit using the screws that were
removed earlier.
Removing the Source Panel for Repolishing (cont.)
Source Panel