7. On the UV Setting screen, by holding the UV 24h
Mode slide key and sliding it right, the UV 24h
Mode is turned to ON. Press Apply key to start the
UV 24-hour mode. The display returns to the Tools
#1 screen.
8. The UV lamp lights continuously for 24 hours. “UV : ON” is displayed on the UV lamp condition display
when UV lamp is lighting.
Note:
The UV 24-hour mode may cause the automatic set temperature alarm because of a rising chamber
temperature.
After procedure 8, by opening the outer door when UV lamp is lighting, UV lamp is turned OFF and UV
24-hour mode is canceled by opening the outer door. Redo from procedure 4 to start the UV 24-hour mode
again.
9. Press Top key to return to the Top screen.
10. 24 hours after, UV lamp turns OFF automatically. Install all attachments removed in the procedure 1.