●Table 7~9 show the behavior of the alarm (buzzer) and Ring Back function when pressing Buzzer key.
Table 7 In the case of other than table 8 or table 9.
Remote Alarm setting
Ring Back
setting
Buzzer from O
2
/CO
2
incubator Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is not
canceled)
ON
ON
ON
(Under continuation)
OFF OFF
OFF: Interlock
with Buzzer key
ON ON
OFF (Alarm is
not canceled)
ON
OFF OFF OFF
Note: Resolve the cause of the alarm in reference to page 91~94 because the alarm itself is not
deactivated by pressing Buzzer key.
Table 8 In the case of high limit temperature alarm or Err38 (The outer door opens during H
2
O
2
decontamination; refer to next page).
Remote Alarm setting
Ring Back
setting
Buzzer from O
2
/CO
2
incubator Remote Alarm
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
ON
ON
(Under continuation)
ON
(Continue)
ON
(Under continuation)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
Note: Close the outer door when Err38 is activated.
Table 9 In the case of Err01 (CO
2
gas cylinder empty), Err02 (N
2
/O
2
gas cylinder empty), Err11, 12
(CO
2
sensor error), Err18 (UV lamp failure) or door alarm*
2
.
Remote Alarm setting
Ring Back
setting
Buzzer from O
2
/CO
2
incubator Remote Alarm setting
When pressing
the Buzzer key
When the Ring Back
set time passes
When pressing
the Buzzer key
When the Ring Back
set time passes
ON: Non-interlock
with Buzzer key
ON
OFF
(Alarm is
canceled)
OFF
(Alarm is already
canceled)
OFF
(Alarm is
canceled*
2
)
OFF
(Alarm is already
canceled*
2
)
OFF
OFF: Interlock
with Buzzer key
ON
OFF
*2: In the door alarm, the remote alarm does not work.
Note:
・When Err01 is activated, connect the new CO
2
gas cylinder and press the Buzzer key to stop the buzzer.
In addition, when the optional MCO-21GC is installed and the gas supply is switched to the reserve gas
cylinder, press the Buzzer key and replace the gas cylinder.
・When Err02 is activated, press the Buzzer key and replace the N
2
gas (O
2
gas) cylinder.