Bulletin TI-FID1000C FID 1000 and 2500 Gas Stations
Parker Hannifin Corporation
Filtration and Separation Division
Haverhill, MA • 1-800-343-4048
www.parker.com/ags
19
1 Turn the Power Switch to OFF and make sure that the customer installed Shut off valve is
in the OPEN position in order to vent the hydrogen from the generator.
2 Lift the top cover lid and unscrew the water reservoir cap. Remove the Resin bed car-
tridge from the Reservoir bottle.
3 Inspect the replacement Resin Bed Deionizer for damage prior to installation into the
water reservoir. Replace the Reservoir cap after the resin bed has been reinstalled.
4 Use the Water Drain Port with the 1/4" tube connected to it to drain the instrument. There
is approximately 2 liters of DIO water in the unit when full.
5 Remove the drain port insert by pressing down on the metal tab and pulling the fitting.
Refill the unit with deionized water by using either the Autofill or by manually filling the
unit with approximately 2 liters of DIO water.
NOTE: Do not allow the DIO Resin bed to dry out; it must be immersed in water to perform.
6 Restart the generator in the same manner found in the Startup section.
Maintenance
Fuse Replacement Occasionally, the fuses in the FID gas station may burn out. The fuses are located in the
power receptacle on the back side of the generator. Before servicing the fuse, turn the unit
off and disconnect the power cord from both the power supply and the generator power
receptacle.
To access the fuses, use a small screwdriver to remove the fuse holder located in the back of
the unit and in the power receptacle. In the FID Gas station, both the phase and the neutral
are fused separately. As a result of this configuration, both fuses should be checked any time
fuse replacement is warranted. Replace either one or both with the appropriate fuse and re-
assemble.
To maintain the safety and performance integrity of the product, use only the fuse of
the size and type detailed in the Specifications, Parts and Accessories sections of this
bulletin.
Don’t Forget To 1 Complete and mail your registration certificate.
2 Keep your product certification in a safe place.
3 Call the Technical Services Department at 800-343-4048, 8am to 5pm Eastern Time with
any questions (North America only). For other locations, please contact your local representa-
tive.
Serial Numbers The serial number for the unit is located on the back of the unit. For your records, and in case
service is required, please record the following:
DATE IN SERVICE______________SERIAL NO_______________
Please have the serial number available when calling for assistance.