14
Physical Size: 5” wide x 9 ¾” high x 12 ¼” deep
Weight: 7 lb.
Fresh Gas Outlet Size: 15mm ID x 22 mm OD
Nitrous Oxide Inlet: Male DISS
Oxygen Inlet: Male DISS
Maximum Oxygen Flow Rate: 10 LPM @ 50-55 PSIG Inlet and 50% setting of mixture dial
Maximum Nitrous Oxide Flow Rate: 10 LPM @ 50-55 PSIG Inlet and 50% setting of mixture dial
Minimum Nitrous Oxide Concentration 0%
Maximum Nitrous Oxide Concentration 70%
Optional Maximum N2O Concentration 50%
Oxygen and Nitrous Oxide Flow Tube
Accuracy (Full Scale) ½ L
Inlet Oxygen Pressure Maximum 50-55 PSIG
Inlet Nitrous Oxide Pressure Maximum 50-55 PSIG
Mixture Dial Calibrations: With supply pressure at 50 to 55 PSIG and total flow of 10 LPM, oxygen
and nitrous oxide flow rates within ½ LPM of mixture dial calibration. (As indicated on individual
flowtubes.)
Total Flow Range 2 to 20 LPM at 50 to 55 PSIG supply pressure and dial mixture set at 50 % (As
indicated on individual flowtubes.)
Total Flow Variation with supply pressure at 50 to 55 PSIG and total flow of 10 LPM, total flow
variation within ½ LPM of total flow throughout mixture dial range. (As indicated on individual
flowtubes.)
B. MDM Portable Stand Mounted Unit (see fig 27)
Maximum Height: 53 inches
Minimum Height: 40 inches
Floor Area: 22 ½ “ sq. Maximum
C. MDM Wall Mounted (see fig 28)
Maximum Distance from Wall: 26 inches, approximately
Minimum Distance from Wall: 5 inches, approximately
Maximum Working Width: 48 ¾ inches, approximately
Maximum Working Height: 22 inches, approximately
D. MDM Cabinet Mounted (see fig 29)
Minimum Cabinet Opening: 6 inches
Minimum Swing Radius: 17 inches
NOTE: All specifications subject to manufacturing tolerance.