EasyManua.ls Logo

Pentair AUTOTROL MAGNUM 742 - Magnum valve cartridge

Pentair AUTOTROL MAGNUM 742
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installer Manual Magnum 742-762 - Troubleshooting
88 / 106 Ref. MKT-IM-012 / C - 25.10.2019
10.3. Magnum valve cartridge
The following procedure provides sequential troubleshooting steps to isolate a suspect cartridge.
Note
The dynamic pressure applied to the valve must be greater than 172.4 kPa at all times when
performing the following tests. If a cartridge needs to be removed, follow the removal
procedure in section 9.6.12. Cartridge replacement, page 83.
Loss of water
pressure.
Fouled resin bed due to iron
accumulation.
Clean Magnum control valve and
mineral bed with cleaner.
Slots in riser pipe or laterals are filled
with resin fines.
Inspect and clean distributor pipe
slots as needed.
Salt in water to
service after
regeneration.
The injector is too small for the system
size.
Install the correct injector.
Brine draw time excessively long due
to low water pressure.
Increase water pressure above
172 kPa minimum.
Restricted drain line.
Remove drain line restriction.
Insufficient rinse volume.
Increase slow rinse time, fast rinse
time or both.
Plugged injector and/or injector
screen.
Inspect and clean the injector and/
or injector screen assembly.
Issue Cause Reset and recovery
#4 Inlet cartridge
#3 Outlet cartridge
#1 Drain cartridge
#2 Rinse cartridge

Table of Contents

Related product manuals