TXP
™
QUICK REFERENCE GUIDE
1. Mount TXP
™
5 to pole.
2. Connect AC power source.
3. Connect gas source.
4. Connect breathing circuit.
5. Always perform a pre-use check before
ventilating any patients.
6. Recommended HME/HMEF bacterial/viral
lter with an eciency
of >99.999% with a
2 cmH
2
O or less
ow resistance
at 60 L/min.
7. Set Frequency arrow straight
up for starting position, which
will produce a high-frequency
rate of approximately 500.
8. Connect to patient invasively or
non-invasively.
9. Turn Amplitude counterclockwise
to start ventilation.
Display will turn on as soon as ventilation pressure
is detected above 2.5 cmH
2
O for 5 seconds.
Alarms default
set 45 and 3
10. Press once for alarm
silence if needed.
Simple System Setup
This device is
NOT an oscillator (HFOV) NOT a jet (HFJV)
NOT a Conventional Ventilator
It is a High-Frequency
Percussive Ventilator (HFPV)
Optional secondary
lters may be added
to the two 15 mm ports,
either directly or with tubing
connectors.
Optional PEEP
valve with or
without lter
Optional
one-way
valve
Intubated patient with Phasitron®
Supplemental oxygen can be
added after lter