CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
26
PERISO SA
If the red light is on, only after having disconnected the device from the mains, open the red cap of
the tank (indicated in the image above) and fill the tank using only demineralised water and the
supplied bottle. Close the cap and reconnect the device to the mains, leaving the handpiece lying
on the floor. Restart the device and let it run for about 10-15 minutes and check that the light is
blue / green again.
Turn off the device, reposition the handpiece on the arm and restart.
Check that the magnetic field output is correct.
Blue / green light: correct water level in the
radiator.
Red light: insufficient water level in the radiator.
It is necessary to fill the tank.