CTU MEGA 20 Diamagnetic Pump MOD07-2-5 Rev.13 of 28/08/2019
6
PERISO SA
Avoid using the device in combination with monitoring devices.
Portable and mobile communications equipment can affect the operation of the
device.
In the diamagnetic pump mode the handpiece can be used between 4/5 cm
away from the skin of the patient.
ATTENTION:
only distilled water should be used in the tank.
Connection to the power supply
Connecting the device to the power supply is simple and immediate, and must be made
by the use of the power cable following the instructions below (Chapter 3).
To avoid the risk of electric shock, this device must be connected only to mains
supplies with a protective ground connection
If the power cable is damaged, it must be replaced only by the manufacturer
Periso or by personnel appropriately qualified by Periso.
Any repair work on the CTU Mega 20 product must be carried out by the
manufacturer Periso or by personnel appropriately qualified by Periso.
The device is not permanently installed and uses the power plug for separation
and isolation from the mains voltage.
The distance between the device and the power cable should be almost equal to
90 cm in order to unplugging the device if needed.
Moving the device
The device is equipped with wheels to facilitate moving in various environments. While
moving the device, it is advisable to hold the device to the outer metal frame making sure
that the wheel brakes are unlocked, as shown in figure.
Check that the wheels are unlocked before moving the device.
Check that the wheel brakes are locked every time you start a treatment.
Make sure before starting a treatment that the device is blocked on a flat floor.
It is forbidden to push the device on its side since it can cause it to tip over.