Maintenance
103
Replacing a Filament
This procedure is the same for both an EI or CI source.
CAUTION
ATTENTION
Make sure you are wearing lint-free, powder-free nitrile, nylon, or
PVC gloves, and that you wipe each part with a methanol
dampened Kimwipe.
Assurez-vous que vous portez des gants en nitrile, en nylon ou en
PVC sans peluche, sans poudre et que vous essuyez chaque partie
avec un Kimwipe humidifié au méthanol.
Items and Tools Required
Filament assembly (Part No. E6470012).
Methanol.
Lint-free, powder-free nitrile, nylon, or PVC gloves.
Aluminum foil or lint-free disposable cloth squares.
To replace a filament, follow this procedure.
1. Prepare a clean, uncluttered work area and place a square of aluminum foil with
the shiny side up.
2. Remove the source by following the procedure Removing and Returning the
Source on page Error! Bookmark not defined..