10 Repair and Disassembly
194
7
Refit the noninvasive blood pressure tubing first (before reconnecting the ribbon cable).
When refitting the tubing, also make sure you connect it to the left-hand connector when facing
the front of the sockets.
8 Then reconnect the ribbon cable. Make sure the cable lock is firmly closed and the cable is secure.
9 Replace the top cover (see “Replacing the Top Cover Assembly” on page 180).