12 Specifications
224
WARNING
Explosion Hazard: Do not use in the presence of flammable anesthetics, such as a flammable
anesthetic mixture with air, oxygen, nitrous oxide, or in oxygen rich environment. Use of the devices in
such an environment may present an explosion hazard.
Physical Specifications
Fetal Monitor
Fetal Monitor Sounds
Tympanic Temperature
Operating Temperature Range 16ºC—33ºC (60.8ºF—91.4ºF)
Monitor Physical Specifications M2702A/M2703A
Power Supply Voltages 100 VAC—240 VAC ±10%
Supply Frequency Range 50 Hz—60 Hz
Power Consumption
(current)
0.7—0.4 A (M2702A/M2703A)
1.3—0.7 A (M8023A#E25
Dimensions and
Weight
Size (without options) mm/
(in):
width x height x depth
286 x 133 x 335 mm ±1%
(11.3 x 5.2 x 13.2 in ±1%)
Weight <5.1 kg (11.2 lbs)
Degree of Protection Against Electrical Shock Type CF
Electrical Class Class II equipment
Electrical Power Source External (AC) without battery option #E25
Internal (LiIo battery) if with battery option #E25
Mode of Operation Continuous operation
Water Ingress Protection Code IP X1 (provided recorder drawer is shut)
Global Speed 6.25 mm/sec, 12.5 mm/sec, 25 mm/sec, 50 mm/sec
Startup Time Time taken from switching
on the monitor to seeing the
first parameter labels
<30 seconds
Source Description
Patient alarms and INOPs See the sections on Patient Alarms and INOPs, Standard Philips Alarms, and
ISO/IEC Standard Audible Alarms in the Instructions for Use.
Ultrasound Doppler Direct transmission of Doppler echoes to the speaker of the fetal monitor.
Pulse from SpO
2
, MECG, DECG QRS tone