12 Specifications
252
Fetal Sensor Socket symbol IP 67 Ingress Protection code according
to IEC 60529. The IUP/ECG
patient module (M2738A) is rated
IP 67 (protection against dust,
access to hazardous parts, and the
effects of continuous immersion
in water to a depth of 0.5 meter
for 30 minutes).
SpO
2
Socket symbol IP 68 Ingress Protection code according
to IEC 60529. All transducers
(excluding M2738A) are rated
IP 68 (protection against dust,
access to hazardous parts, and the
effects of continuous immersion
in water to a depth of 1.0 meter
for five hours).
Noninvasive Blood Pressure
Socket symbol
IP X1 Ingress Protection code according
to IEC 60529. The monitors and
interface cable for the Avalon CTS
(M2731-60001/M2732-60001) are
rated IP X1 (protection against
water dripping vertically only).
Type BF IP 31 Ingress Protection code according
to IEC 60529 (protection against
condensation only).
Symbol indicating the monitor
has the triplets option
IP 21 Ingress Protection code according
to IEC 60529 (protection against
ingress of water when the water is
dripping vertically)
Symbol indicating the monitor is
capable of intrapartum
monitoring
Type CF equipment, not
defibrillation proof
Button to open paper drawer/
paper eject. (FM40/FM50)
Indicates location of service
number
Protective earth terminal (FM40/
FM50)
Indicates location of serial number
Equipotential grounding point
(FM40/FM50)
Indicates location of catalog
number