Accessing the Diagnostics Menu and Maintenance Options
Performance Verification Testing
3-6 Efficia CM Series Service Guide
Accessing the Diagnostics Menu and Maintenance Options
Some of the verification tests require you to use the Diagnostics Menu or the
Maintenance options.
After you access the Diagnostics Menu, you must enter an additional password to
access the Maintenance options. Only trained biomedical engineers or support
personnel should access the Maintenance options and perform maintenance and repair
procedures.
Caution When you open the Diagnostics Menu, monitoring is suspended and measurements
stop.
3
Enter the current Administrator password, and then select OK.
Note — The default Administrator password is 215. The password may
have been changed for security purposes in your facility.
4
In the Enter Password window, select OK.
The Admin tab opens.
Caution
The password-protected tabs remain unlocked for 1 minute after you
close them. This allows you to re-open the tab without having to
re-enter the password. Do not leave the monitor unattended during
the unlock time.
The following options are described in this guide. All other options are
described in the Efficia CM Series Patient Monitors Installation and
Configuration Guide.
Demo ModeDemo Mode allows you to demonstrate the monitor
without actually monitoring parameters.
Upgrade SoftwareOpens the Upgrade Software menu. For more
information, see “Software Upgrades” on page 2-9.
DiagnosticsOpens the Diagnostics Menu. Monitoring is suspended
while this menu is open. For more information, see “Accessing the
Diagnostics Menu and Maintenance Options” on page 3-6.
Note — The Diagnostics button is unavailable when the monitor is in
Demo mode.
View Logs — Opens the View Log menu, which allows viewing and
management of system logs. For more information, see “Managing
Logs” on page 4-20