Specifications
Efficia CM Series Rack and Modules Instructions for Use 9-15
Alarm Specifications
CO
2
Check Airway Adapter High CO
2
Adapter is not plugged in
correctly or has been removed.
There is an optical blockage on the
windows of the airway adapter.
May also be caused by a failure to
perform a zero when the adapter
type is changed.
To clear, clean the airway adapter
if mucus or moisture is seen. If the
adapter is clean, perform a zero.
CO
2
Sensor Over Temp Low CO
2
Sensor temperature is greater
than 40° C.
CO
2
Equip Malfunc Low CO
2
Equipment Malfunction or if
the module configuration is not
successful. Check error log for
details.
CO
2
Purging Low The CO
2
filter line is being purged
to remove an occlusion in the line
or airway adapter. If the occlusion
is removed, the message
disappears.
1. Also displayed in the Error log
Alarm Message Priority Cause