8: Pacing Pacing View
92
Pacing View
Pacing View appears when the Therapy knob is turned to the Pacer position. Pacing View includes a
status block which appears in Wave Sector 3 of the display (see Figure 53). The patient's internal
paced status is not displayed. Pacing-related information in Pacing View includes:
•Pacing Markers: Markers, indicating a pace pulse was delivered, appear in Wave Sector 1 (and
in Wave Sector 2, if the wave is cascading) each time a pacer pulse is delivered.
•R-Wave Arrows: R-Wave arrows appear in Wave Sector 1 (and in Wave Sector 2, if the wave is
cascading) when in Demand Mode pacing. R-Wave arrows do not appear on paced beats.
•Pacing Status: Indicates the current pacing status.
• When pacing is active,
is displayed when the device is on AC power.
• If the device is running on battery,
is displayed.
• If pacing is not active,
is displayed.
• Pacing Alarm: If there is a pacing-related alarm during pacing, the current pacing status is
replaced with an alarm message. See “Pacing Alarms” on page 97.
• Pacing Mode: Indicates if the device is in Demand or Fixed Mode Pacing.
• Pacing Rate: Indicates the current pacing rate, including unit of measure.
• Pacing Output: Indicates the current output, including unit of measure.
• Alarms: Are on automatically.
WARNING: If pacing is interrupted for any reason, you must press the soft key to resume pacing.
Figure 53 Pacing View Layout
Pacing
Markers
Pacing Bar:
Mode,
Status,
Alarms
Pacing
Rate Value
Pacing
Output Value
R-Wave
Arrows
Pacing Control soft key: Press
to pause pacing; press to start or
resume pacing
Pacer Settings soft key: Press to
bring up the menu for adjusting Pacer Rate and
Output.